A ResMed AirSense 11 at its published typical power consumption of 56.1 W uses about 56.1 W x 8 hours = 450 Wh in a night, so a 1,024 Wh power station covers about 1,024 x 0.85 / 450 = 1.9 nights. Turning the heated humidifier and heated tube off lowers the draw a great deal, which is why the same station can stretch to several nights for some users.
This page is energy math, not medical advice, and a portable power station is not a medical backup system. Before you plan backup power for a CPAP, ask the CPAP maker or your equipment supplier which power sources the machine supports, and ask your doctor or supplier what backup arrangement suits you. Talk with your doctor before changing any therapy setting, including humidification.
Check the power station maker’s guidance too. Jackery, for example, states on its website that its power stations are not recommended for use with medical devices; check your own brand’s manual and terms for similar language. If either maker advises against the pairing, follow that advice; the runtime figures below only describe how much energy a station holds.
The CPAP Runtime Formula
Nights of runtime = rated Wh x 0.85 / (average CPAP watts x hours of sleep). The 0.85 is a common planning figure for inverter loss when the CPAP’s power supply plugs into the station’s AC outlet.
- Average CPAP watts: the AirSense 11 user guide lists 56.1 W typical and 73.2 W peak power consumption, with a 65 W power supply. Your average depends on pressure, humidifier level, heated tube temperature, and room temperature.
- Hours of sleep: use your actual nightly therapy hours from the device or its app. This guide uses 8.
- Rated Wh: the station’s capacity from its spec sheet.
Humidifier On vs Off: The Math
Heat is the expensive part of the energy budget. ResMed’s own description of its Power Station II battery, sold for earlier AirSense 10 and S9 machines, puts runtime at roughly 10 to 14 hours without the humidifier and heated tube, and about 2 to 3 hours with both on. That is a drop of roughly 4 to 5 times from the same battery.
ResMed does not publish a humidifier-off wattage for the AirSense 11 in the user guide, so the second row below uses 20 W as a labeled example. Measure your own machine with a plug-in watt meter over a full night and replace it.
| Setting | Average draw | Wh per 8 h night | Rated Wh needed per night (/ 0.85) |
|---|---|---|---|
| Humidifier and heated tube on (ResMed typical figure) | 56.1 W | 56.1 x 8 = about 450 Wh | about 530 Wh |
| Humidifier and heated tube off (example) | 20 W | 20 x 8 = 160 Wh | about 190 Wh |
Nights per Station Size
Capacities are manufacturer ratings. Nights are usable Wh / Wh per night, rounded down.
| Station (rated / usable) | Heat on, 450 Wh a night | Heat off, 160 Wh a night (example) |
|---|---|---|
| 300 Wh class (300 / 255 Wh) | about 4.5 hours, not a full night | about 1.5 nights |
| Jackery Explorer 1000 v2 (1,070 / 910 Wh) | about 2 nights | about 5.5 nights |
| EcoFlow DELTA 3 Max (2,048 / 1,740 Wh) | about 3.8 nights | about 10.8 nights |
The 300 Wh row answers a common search. A “300 W” station usually refers to output, not capacity, so check the Wh figure; for a unit that holds about 300 Wh, 255 / 56.1 = about 4.5 hours with heat on.
AC Outlet or DC Cable
The AirSense 11 runs on 24 V DC from its power supply, and ResMed lists an Air11 DC/DC converter among its accessories. A DC route skips the station’s inverter, so less energy is lost than on the AC outlet. Whether a given converter and station work together is a question for ResMed or your equipment supplier; use only connection methods the device maker supports.
A Winter Outage Scenario
Room temperature is part of the plan. The AirSense 11 user guide lists an operating range of 41°F to 95°F, and a bedroom in an unheated house can fall below that during a long winter outage. A two-night winter plan with heat on needs about 2 x 450 / 0.85 = about 1,060 Wh rated, plus whatever the household needs for lights and a phone, which points to the 2,000 Wh class. Ideas for holding a room’s temperature without grid power are in how to keep a house warm without power.
Common Mistakes
- Planning with the peak figure only. 73.2 W x 8 h = about 586 Wh gives a conservative ceiling, but it can make a right-sized station look too small. Use your measured average and keep the peak as a margin check.
- Forgetting the rest of the house. If the same station also runs a fridge, add its Wh; see how long a power station will run a fridge.
- Leaving the station at 50 percent. Keep it topped up between outages so the first night starts from full.
- Skipping the maker’s guidance. Ask ResMed, Philips, or your supplier whether the station’s output type suits the device; background on waveform is in pure sine wave inverters explained.
For general runtime math beyond CPAP, see how long a portable power station will run. For a shortlist of stations by capacity and output, see the best portable power stations for home backup.
FAQ
How much power does a CPAP use?
ResMed lists 56.1 W typical and 73.2 W peak for the AirSense 11. Other machines differ, so read your device’s user guide or measure it with a watt meter.
How long will a CPAP run on battery?
Usable Wh / average watts. On a station with about 910 usable Wh, that is 910 / 56.1 = about 16 hours with heat at the typical figure, and much longer with heat off.
Should I turn off the humidifier to save battery?
That is a therapy decision, not a battery decision. The math shows how much energy it saves; ask your doctor or equipment provider whether it suits you.
Changelog: November 2026: first published.